H2S-625 H2S Analyzer
ATEX and IECEx Certified for Hazardous Areas

Configurazioni Opzionali:
- Sample System Add-On Components
- Bi-Directional RS485 Modbus RTU
- Involucro per condizioni climatiche estreme
- Extended Temperature Ranges
Applicazioni:
- Estrazione di Gas Naturale e Gasdotti
- Trattamento del Gas Naturale
- Monitoraggio del Gas di Discarica
- Biogas prima e dopo lo scrubber HโS
- E Molte Altre Applicazioni Industriali
- ATEX / IECEx Certified for Hazardous-Area Applications
- Designed for Natural Gas, Biogas & Industrial HโS Applications
- Campi di misura da 0โ10 PPM a 0โ10.000 PPM HโS
- Tecnologia di sensore elettrochimico HโS di precisione
- Large Backlit Display with Intuitive Menu Interface
- Two Fully Configurable Alarm Relay Contacts
- Optional MODBUS RS485 RTU Communications
Specifiche Tecniche:
- Precisione: < ยฑ1% del fondo scala
- Allarmi: 2 Contatti Relรจ Configurabili
- Analyzer Range: Multiple up to 0 - 10000 ppm
- Area Classification: Zone 1 IIB + H2
- Dimensioni: 12" ร 12" ร 5,25"
- Flow: 0.50 - 5.0 SCFH
- Gas Connections: 1/4โ Compression Tube
- Uscita: Isolata 4โ20 mA oppure 1โ5 VDC
- Alimentazione: 12โ24 V CC, 100โ240 V CA
- Pressione: ingresso 0,1โ50 PSIG, sfiato in atmosfera
- Tempo di risposta: T90 in 45 secondi
- Sistema di campionamento: opzionale
- Sensore: cella elettrochimica di precisione
- Temperatura: 0โ50 ยฐC
- Garanzia sensore: 12 mesi
- Garanzia elettronica: 12 mesi
- Peso: 21 lb
Sensibilitร incrociata
- Risposta sensore per 100 ppm di gas presente:
- Metilmercaptano: 40 ppm
- Monossido di carbonio: 4 ppm
- Idrogeno: 1 ppm
- Biossido di zolfo: 18 ppm
H2S-625 H2S Analyzer | Product Specifications
Analizzatore HโS:
The H2S-625 Hydrogen Sulfide Analyzer combines a rugged hazardous-area design with Southland Sensingโs precision electrochemical HโS sensor technology. Designed for continuous monitoring in natural gas, biogas, and other demanding industrial applications, the analyzer provides reliable HโS measurement in a compact, easy-to-operate configuration.
The H2S-625 is certified for installation in Zone 1, Group IIB + Hโ hazardous-area environments, making it suitable for applications where third-party certified instrumentation is required.
Both the power input and analog output are electrically isolated, helping prevent ground loops and other electrical interference that can affect signal stability when the analyzer is integrated with plant control and monitoring systems.
Process gas connections are provided through 1/4″ compression tube fittings. The integral sample system includes a Sample/Span selection valve, flow control, and flow indication for straightforward operation, calibration, and maintenance.
Optional bi-directional MODBUS RS485 RTU communications are available for integration with plant control, SCADA, and data acquisition systems.
Tecnologia sensore HโS:
The HโS sensor used in the H2S-625 is based on proven electrochemical fuel cell technology and is designed for reliable hydrogen sulfide measurement across a wide range of industrial applications. Southland Sensing HโS sensors are manufactured under a controlled quality program to provide consistent performance, accuracy, and long-term stability.
Il sensore autonomo richiede minima manutenzione periodica, senza necessitร di pulire elettrodi o reintegrare elettrolita.
La tecnologia dei sensori elettrochimici HโS di precisione Southland Sensing รจ progettata per misure affidabili, maggiore vita utile del sensore e minori esigenze di manutenzione.
Requisiti di Alimentazione:
Input Power: 100โ240 VAC, 50/60 Hz or 12โ24 VDC
DC Current Draw: 50 mA nominal (DC operation)
Sensori di ossigeno:
H2S-1x PPM H2S Sensor: 0 – 200 Max PPM Range
H2S-2x PPM H2S Sensor: 0 – 2000 Max PPM Range
H2S-3x PPM H2S Sensor: 0 – 10000 MAX PPM Range
I sensori HโS devono essere calibrati periodicamente. Il produttore raccomanda ogni 1โ3 mesi oppure secondo lโapplicazione. La calibrazione con gas campione certificato garantisce unโeccellente linearitร .