OMD-640 Oxygen Analyzer
Portable PPM Oxygen Analyzer with USB Data Logging & Sample System

Technische Daten:
- Genauigkeit: < ±1 % des Messbereichsendwerts*
- Analysis Range: 0-1/10/100/1000ppm/25%
- Batterieanzeige: in die große Anzeige integriert
- Kalibrierung: regelmäßig
- Datenaufzeichnung: herausnehmbarer USB-Stick
- Abmessungen: 10,9 × 10,0 × 4,9 Zoll
- Anzeige: groß, mit Hintergrundbeleuchtung
- Gehäuse: gebürsteter Edelstahl
- Flow Sensitivity: 0.5 - 5 SCFH
- Gasanschlüsse: 1/8-Zoll-Klemmringrohr
- Analogausgang: 0–1 V DC
- Power: Rechargeable Battery 100 - 240 VAC AC Adapter
- Druck: Eingang 0–50 PSIG
- Response Time: T90 in 7 Seconds
- Sample System: Flow Control, 4-way sample/ Bypass Valve, Flow Indicator
- Sensor: TO2-133 Trace O2 Sensor
- Sensor Life: 20 - 25 months
- Temperatur: 0–50 °C
- Temperaturkompensation: integriert
- Garantie: 12 Monate
- Weight: 11.75 lbs
- 0–1 PPM O₂ Low Range with PPB Resolution
- Große hintergrundbeleuchtete Anzeige mit intuitiver Menüführung
- Datenaufzeichnung auf herausnehmbarem USB-Stick
- Integral Bypass / Isolation System
- Kontinuierliche Analyse während des Ladevorgangs
- Präzise elektrochemische Sauerstoffsensortechnik
The OMD-640 combines a rugged portable design with Southland Sensing’s precision electrochemical oxygen sensor technology for demanding trace and ultra-trace oxygen measurements. Designed for field and plant use, the analyzer provides reliable measurement from the PPM range into the sub-PPM region with a straightforward, easy-to-use interface.
The analyzer features a 0–1 PPM O₂ full-scale low range, providing the sensitivity required for low-PPM and sub-PPM oxygen measurements. A removable USB flash drive provides convenient data logging in .CSV format, allowing recorded measurements to be easily transferred to a computer for review, documentation, or further analysis.
The high-visibility display is designed for readability in direct sunlight, making the OMD-640 well suited for outdoor and field applications. An integral Swagelok® bypass valve sample system allows the analyzer to be purged efficiently and helps minimize response time when making trace oxygen measurements.
The OMD-640 utilizes Southland Sensing’s galvanic electrochemical oxygen sensor technology. All oxygen sensors are manufactured in-house under a controlled quality program and are self-contained, requiring minimal routine maintenance with no electrodes to clean or electrolyte to replenish.
Two sensor chemistries are available to accommodate different process gas compositions. The standard TO2-133 sensor is designed for applications containing background gases such as hydrogen, helium, and hydrocarbons. For applications containing acid gases, including carbon dioxide and many natural gas streams, the TO2-233 sensor provides an acid-resistant alternative.
The combination of portable operation, trace oxygen sensitivity, integral bypass sample system, field data logging, and application-specific sensor options makes the OMD-640 a versatile solution for demanding trace and ultra-trace oxygen measurements.
Anwendungen:
- Pipeline leak detection
- Spot Checking Air Separation & Liquification
- Headspace Gas Analysis
- Beverage Grade CO2 Monitoring
- Heat Treating & Bright Annealing
- Inert Gas Welding of Exotic Materials
Optionales Zubehör
- ENC-640 Tragekoffer mit Schaumstoffeinlage
- CF-640 Coalescing filter with 0.1u filter
- PP-640 integrierte Probenahmepumpe
- TO2-233 Oxygen Sensor (> 0.5% CO2 present)
- ENC-640 Tragekoffer mit Schaumstoffeinlage
- PP-640 integrierte Probenahmepumpe