OMD-640 Oxygen Analyzer
Portable PPM Oxygen Analyzer with USB Data Logging & Sample System

Spécifications :
- Précision : < ±1 % de la pleine échelle*
- Analysis Range: 0-1/10/100/1000ppm/25%
- Indicateur de batterie : intégré au grand écran
- Étalonnage : périodique
- Enregistrement de données : clé USB amovible
- Dimensions : 10,9 × 10,0 × 4,9 pouces
- Affichage : grand écran rétroéclairé
- Boîtier : acier inoxydable brossé
- Flow Sensitivity: 0.5 - 5 SCFH
- Raccord gaz : tube à compression 1/8 pouce
- Sortie analogique : 0–1 V CC
- Power: Rechargeable Battery 100 - 240 VAC AC Adapter
- Pression : entrée 0–50 PSIG
- Response Time: T90 in 7 Seconds
- Sample System: Flow Control, 4-way sample/ Bypass Valve, Flow Indicator
- Sensor: TO2-133 Trace O2 Sensor
- Sensor Life: 20 - 25 months
- Température : 0–50 °C
- Compensation de température : intégrée
- Garantie : 12 mois
- Weight: 11.75 lbs
- 0–1 PPM O₂ Low Range with PPB Resolution
- Grand écran rétroéclairé avec menus intuitifs
- Enregistrement sur clé USB amovible
- Integral Bypass / Isolation System
- Analyse continue pendant la recharge
- Precision Electrochemical Oxygen Sensor Technology
The OMD-640 combines a rugged portable design with Southland Sensing’s precision electrochemical oxygen sensor technology for demanding trace and ultra-trace oxygen measurements. Designed for field and plant use, the analyzer provides reliable measurement from the PPM range into the sub-PPM region with a straightforward, easy-to-use interface.
The analyzer features a 0–1 PPM O₂ full-scale low range, providing the sensitivity required for low-PPM and sub-PPM oxygen measurements. A removable USB flash drive provides convenient data logging in .CSV format, allowing recorded measurements to be easily transferred to a computer for review, documentation, or further analysis.
The high-visibility display is designed for readability in direct sunlight, making the OMD-640 well suited for outdoor and field applications. An integral Swagelok® bypass valve sample system allows the analyzer to be purged efficiently and helps minimize response time when making trace oxygen measurements.
The OMD-640 utilizes Southland Sensing’s galvanic electrochemical oxygen sensor technology. All oxygen sensors are manufactured in-house under a controlled quality program and are self-contained, requiring minimal routine maintenance with no electrodes to clean or electrolyte to replenish.
Two sensor chemistries are available to accommodate different process gas compositions. The standard TO2-133 sensor is designed for applications containing background gases such as hydrogen, helium, and hydrocarbons. For applications containing acid gases, including carbon dioxide and many natural gas streams, the TO2-233 sensor provides an acid-resistant alternative.
The combination of portable operation, trace oxygen sensitivity, integral bypass sample system, field data logging, and application-specific sensor options makes the OMD-640 a versatile solution for demanding trace and ultra-trace oxygen measurements.
Applications :
- Pipeline leak detection
- Spot Checking Air Separation & Liquification
- Headspace Gas Analysis
- Beverage Grade CO2 Monitoring
- Heat Treating & Bright Annealing
- Inert Gas Welding of Exotic Materials
Accessoires en option
- Mallette ENC-640 avec mousse de protection
- CF-640 Coalescing filter with 0.1u filter
- Pompe d’échantillonnage intégrée PP-640
- TO2-233 Oxygen Sensor (> 0.5% CO2 present)
- Mallette ENC-640 avec mousse de protection
- Pompe d’échantillonnage intégrée PP-640