OMD-640 Oxygen Analyzer
Portable PPM Oxygen Analyzer with USB Data Logging & Sample System

技术规格
- 精度:< 满量程的 ±1%*
- Analysis Range: 0-1/10/100/1000ppm/25%
- 电池指示:集成于大显示屏
- 校准:定期
- 数据记录:可拆卸 USB 闪存盘
- 尺寸:10.9 × 10.0 × 4.9 英寸
- 显示屏:大屏幕,带背光
- 外壳:拉丝不锈钢
- Flow Sensitivity: 0.5 - 5 SCFH
- 气体连接:1/8 英寸压缩管
- 模拟输出:0–1 V DC
- Power: Rechargeable Battery 100 - 240 VAC AC Adapter
- 压力:入口 0–50 PSIG
- Response Time: T90 in 7 Seconds
- Sample System: Flow Control, 4-way sample/ Bypass Valve, Flow Indicator
- Sensor: TO2-133 Trace O2 Sensor
- Sensor Life: 20 - 25 months
- 温度:0–50 °C
- 温度补偿:内置
- 保修:12 个月
- Weight: 11.75 lbs
- 0–1 PPM O₂ Low Range with PPB Resolution
- 大型背光显示屏,菜单操作直观
- 通过可拆卸 USB 闪存盘记录数据
- Integral Bypass / Isolation System
- 充电期间可持续分析
- Precision Electrochemical Oxygen Sensor Technology
The OMD-640 combines a rugged portable design with Southland Sensing’s precision electrochemical oxygen sensor technology for demanding trace and ultra-trace oxygen measurements. Designed for field and plant use, the analyzer provides reliable measurement from the PPM range into the sub-PPM region with a straightforward, easy-to-use interface.
The analyzer features a 0–1 PPM O₂ full-scale low range, providing the sensitivity required for low-PPM and sub-PPM oxygen measurements. A removable USB flash drive provides convenient data logging in .CSV format, allowing recorded measurements to be easily transferred to a computer for review, documentation, or further analysis.
The high-visibility display is designed for readability in direct sunlight, making the OMD-640 well suited for outdoor and field applications. An integral Swagelok® bypass valve sample system allows the analyzer to be purged efficiently and helps minimize response time when making trace oxygen measurements.
The OMD-640 utilizes Southland Sensing’s galvanic electrochemical oxygen sensor technology. All oxygen sensors are manufactured in-house under a controlled quality program and are self-contained, requiring minimal routine maintenance with no electrodes to clean or electrolyte to replenish.
Two sensor chemistries are available to accommodate different process gas compositions. The standard TO2-133 sensor is designed for applications containing background gases such as hydrogen, helium, and hydrocarbons. For applications containing acid gases, including carbon dioxide and many natural gas streams, the TO2-233 sensor provides an acid-resistant alternative.
The combination of portable operation, trace oxygen sensitivity, integral bypass sample system, field data logging, and application-specific sensor options makes the OMD-640 a versatile solution for demanding trace and ultra-trace oxygen measurements.
应用领域
- Pipeline leak detection
- Spot Checking Air Separation & Liquification
- Headspace Gas Analysis
- Beverage Grade CO2 Monitoring
- Heat Treating & Bright Annealing
- Inert Gas Welding of Exotic Materials
可选配件
- ENC-640 带泡沫内衬的便携箱
- CF-640 Coalescing filter with 0.1u filter
- PP-640 内置采样泵
- TO2-233 Oxygen Sensor (> 0.5% CO2 present)
- ENC-640 带泡沫内衬的便携箱
- PP-640 内置采样泵