Glossary

Also referred to as a vocabulary or definition of terms, we have put together some common terms that come up in reference to our precision O2 sensors.  If you run across a term you are not familiar with, contact us.

  • ACCURACY:  The quality of closeness to a specified value under stated conditions. Usually expressed as uncertainty – the maximum deviation to be expected between a reading or output and the actual value being measured.
  • AMBIENT AIR:  Often used as a term when referencing oxygen analyzer calibration.  Ambient air consists of 20.9% oxygen which can be used as a known SPAN gas value.
  • AMBIENT TEMPERATURE:  The minimum / maximum temperature of air surrounding the device or equipment.  A critical parameter when specifying an analyzer.
  • ANALOG:
    • A parameter which varies in a continuous, rather than incremental or discrete-step manner.
  • CALIBRATE:
    • To determine the indication or output of a device with respect to a standard.
    • 0 through 9, a 1/2 digit will display a 0 or 1 and overrange at 2, a 3/4 digit will display up to 3 and overrange at 4. A meter with a display span of ±3999 counts is a 33/4 digit meter.
  • EMI / RFI:
    • Electromagnetic interference / Radio Frequency Interference. 
  • EXPANDED SCALE:
    • An arrangement that expands a specific portion of an overall range to occupy a larger portion of the full-scale length than it normally would.
  • EXPLOSION PROOF:
    • An enclosure capable of withstanding an explosion of a specified gas or vapor which may occur within it and of preventing the ignition of a specified gas or vapor surrounding the enclosures by sparks, flashes, heat or explosion of the gas vapor within.
  • FORM:
    • A method for specifying relay contact configuration: Form A SPST-NO single pole, single throw, normally open Form B SPST-NC single pole, single throw, normally closed Form C SPDT single pole, double throw Change-Over CO same as SPDT
  • FULL SCALE VALUE:
    • The arithmetic sum of the two end-scale values (may not apply to some specialized meters, such as power factor). When zero is not on the scale, the full-scale value is the higher end-scale value. Examples: Scale Full Scale Value 0-10 ppm, 0 – 100ppm, 0 – 500ppm, 0 – 1000ppm, 0 – 2500ppm, 0 – 10000ppm, 0 – 1%, 0 – 5%, 0 – 10%, 0 – 25%
  • GROUND:
    • Reference point for an electrical system. Often used to indicate an earth connection or negative side of a DC supply.
  • LINEARITY ERROR:
    • A measure of the departure from a straight-line response in the relationship between two quantities, where the change in one quantity is directly proportional to a change in the other quantity. Usually expressed as a maximum percent.
  • NEMA 4X / IP66 ENCLOSURE:
    • A rating standard which defines intended use and degree of environmental protection.  Typically associated with analyzers that will be installed outdoors.
  • PPM (PARTS PER MILLION):
    • A convenient format to express very small numbers, such as temperature coefficients. 100ppm is 0.01%. PPM is also written as “trace” as in trace oxygen.
  • PSIA (PSI Absolute):
    • A pressure reading using vacuum as the reference.
  • PSIG (PSI Gauge):
    • A pressure reading using ambient air pressure as the reference.
  • RANGE:
    • The span of values over which a meter will function without entering overload condition, e.g. 0-100ppm
  • REPEATABILITY:
    • The ability of an instrument to register the same reading in successive measurements of the same input.
  • RESOLUTION:
    • The degree to which nearly equal values of a quantity can be discriminated. In digital meters, the value represented by a one-digit change in the least-significant digit.
  • SENSITIVITY:
    • The minimum change in input to which a device can respond.
  • SHIELD:
    • A protective conductive covering that provides a least resistance path to ground for external interference.
  • SPAN CAL ADJUSTMENT:
    • The ability to adjust the gain of an instrument so the display value.
  • T90 Response Time:
    • The time required for a sensor to respond to 90% of its total change resulting from a step input.
  • TWO-WIRE TRANSMITTER:
    • An instrument configuration in which the signal output and power input share two wires, power is transfer on one wire and returned to a PLC or DCS system via a 4 – 20mA signal on the other wire.  In most cases, this signal is non-isolated. 
  • ZERO CAL ADJUSTMENT:
    • The ability to adjust an instrument so that zero output or zero display corresponds to a specific input signal, such as 0V or 4mA.