OMD-677 Oxygen Analyzer
Online Trace Oxygen Analyzer w/ Integral Sample System

The OMD-677 is designed to measure trace oxygen in the sub 1 ppm region. The unit combines an advanced set of electronic features with our, industry leading, precision electrochemical oxygen sensor technology. The result is a highly reliable and cost effective design with easy-to-use user interface.
The analyzer comes with a 0 -1 PPM full scale low range with a resolution of 0.001 ppm. The analyzer can also be configured for 0 – 1 ppm, 0 – 10ppm, 0 – 100ppm, 0 – 1000ppm and 0 – 25% in autorange or manual-range mode.
The analyzer offers the user 2 different digital communication options, both of which are bi-directional. This comes in the form of MODBUS RS485 ASCII or RS232.
Alarm functionality comes in the way of 2 fully adjustable form C non-latching relay contacts. These can be configured as NO or NC and can be set as HIGH or LOW with optional delay mode. A power failure alarm is also standard and comes as a form C nonlatching relay contact.
The oxygen sensor used in the OMD-677 is based on the galvanic electrochemical fuel cell principal. All oxygen sensors are manufactured in house by Southland Sensing Ltd. under a strict quality program.
The standard cell (TO2-133T) is unaffected by other background gases such as H2, He, or Hydrocarbons. The optional acidic cell (TO2-233T) works well when acid gases such as CO2 or Natural Gas are present.
Die Sensoren sind vollständig gekapselt und nahezu wartungsfrei – eine Reinigung der Elektroden oder das Nachfüllen von Elektrolyt ist nicht erforderlich.
Anwendungen:
- Air Liquefaction and Separation
- Pure, Gaseous Hydrocarbon Stream Monitoring
- Gas Purity Certification
- Process Monitoring of Gaseous Monomers
- Halbleiterfertigung
- Protective Atmosphere Blanketing of Feedstock
- 0 - 1 PPM Low Range; 0 - 25% High Range
- Resolution of 0.001 Parts-Per-Million Oxygen
- Integral Sample System w/ Bypass Valve
- Bi-Directional Communication Link
- Extensive Electronic Features & Outputs
- Panel-Montage Konfiguration
- Electrochemical Sensor Technology
Technische Daten:
- Accuracy: < +/- 1% of Full Scale Range
- Alarms: (2) Adjustable Relay Contacts (1) Power Fail Relay Contact
- Analysis Range: 0-1/10/100/1000ppm/25%
- Calibration: Periodically
- Communication: Bi-Directional RS232 -- MODBUS RS485 ASCII
- Dimensions: 10.78 x 7.47 x 7.75 inch
- Display: Large with Backlight
- Enclosure: Brushed Stainless Steel
- Flow Sensitivity: 0.5 - 5 SCFH
- Gas Connections: 1/8” Compression Tube
- Output (Analog): 0 - 1V DC, 0 - 10V DC -- Isolated 4 - 20mA
- Power: 100 - 240 VAC
- Pressure: Inlet, 0 - 50 PSIG
- Range ID: 4 - 20 mA analog output -- 0 - 1 V DC analog output
- Response Time: T90 in 10 Seconds
- Sample System: Flow Control, 4-way sample / Bypass Valve, Flow Indicator
- Sensor: TO2-133T Trace O2 Sensor
- Sensor Life: 20 - 25 months
- Temperature: 0 - 50 deg C
- Temperature Compensation: Digital
- Warranty: 12 months Analyzer & Sensor
- Weight: 12.0 lbs
Direct Fit Field Replacement:
- Designed to replace competitive analyzers:
- Teledyne: 3000TA, 3000TA-XL
- AII: GPR-1600, GPR-1600MS, PI2-MS-1000