How to Measure the Output of an Electrochemical Oxygen Sensor Using a Digital Multimeter
Learn how to measure the output of an electrochemical oxygen sensor in ambient air using a digital multimeter, verify proper sensor output, and compare the measured output against the specification for your sensor model.
Tools Needed
- Multimeter capable of measuring micro-Amp (µA) DC current (i.e. Fluke 87 III)
- Test leads
Step 1: Set Up and Multimeter Configuration
If you are not getting a reading or are experiencing inconsistencies of your electrochemical oxygen sensor, first ensure your testing equipment is set up correctly:
- Confirm your multimeter is powered on and has sufficient battery life.
- Check that the selection dial is set to the current measurement (µA) setting.
- Check that you are reading DC current; some multimeters are preset to AC.
- Ensure the positive (red) test lead is firmly plugged into the current terminal (µA).
- Ensure the negative (black) test lead is firmly plugged into the COM terminal.
Step 2: Checking Output Reading on Sensor
- Verify the sensor is placed on a flat surface with the PCB side facing up.
- Ensure any shorting tab or residue is wiped off the PCB board to avoid shorting.
- Ensure the negative (black) lead is making direct, steady contact with the middle sensor contact.
- Ensure the positive (red) lead is making direct, steady contact with the outer sensor contact.
Step 2A: 0.00 Output Reading
Step 2B: Negative Output Reading
If your multimeter displays a negative value, the polarity of your test leads is probably reversed.
- Remove both test leads from the sensor contacts.
- Reconnect the negative (black) lead specifically to the middle contact.
- Reconnect the positive (red) lead specifically to the outer contact.
Step 2C: Unstable Output Reading
Step 2D: Addressing Out-of-Specification Readings
Step 3: Safe Storage Practices
- Place a red shorting tab (or any shorting tab) across the PCB; make sure that it touches both the inner (negative) and outer (positive) contacts. This will short out the open circuit.
- Store the sensor in an appropriate low-ingress or antistatic bag. Purge the bag slowly with nitrogen or another inert gas and seal the bag.
Output Specifications by Sensor Model
| Class | Sensor Type | Output Specs |
|---|---|---|
| Следовые (PPM) | TO2-1x | 285 – 590 µA |
| Следовые (PPM) | TO2-2x | 285 – 590 µA |
| Следовые (PPM) | TO2-133 | 520 – 1020 µA |
| Следовые (PPM) | TO2-233 | 530 – 1050 µA |
| Следовые (PPM) | TO2-238 | 530 – 1050 µA |
| Гибридный | TO2-19 | 285 – 590 µA |
| Процент | PO2-1x | 245 – 430 µA |
| Процент | PO2-160 | 38 – 74 µA |
| Процент | PO2-24 | 38 – 74 µA |
| Чистота | PO2-1120 | 38 – 74 µA |
Why Measure Sensor Output?
Important Notes:
- Output specifications are based on ambient air (20.9% oxygen), sea level, and 25°C.
- Sensors tested at higher elevations will produce proportionally lower output because the oxygen partial pressure is lower.
- Allow the sensor to stabilize before recording the final reading (typically 45 – 60 seconds).
- Always compare the measured output to the published specification for your sensor model.
Часто задаваемые вопросы
The sensor has likely reached end of life or has been damaged. Alternative check your meter and make sure it is hooked up and functioning properly.
The meter leads are most likely reversed. Connect the black lead to the center contact and the red lead to the outer contact.
Poor electrical contact, a damaged sensor, or insufficient stabilization time may cause fluctuating readings. Often if a sensor has seen a pressure issue outside of the specifications (vacuum or excess pressure) it can damage the membrane and make the sensor unstable.
No. These sensors should be checked by measuring their output current (µA DC).