Descrição
The OMD-507 Online Oxygen Analyzer is a compact, highly configurable oxygen analyzer designed for continuous measurement from trace oxygen levels as low as 0.01 ppm through 100% oxygen. Combining Southland Sensing’s precision electrochemical oxygen sensor technology with an advanced digital electronics platform, the OMD-507 provides reliable oxygen measurement in a rugged, cost-effective package.
An integrated smart MENU interface allows operators to configure, calibrate, and troubleshoot the analyzer directly from the front panel. No laptop, external programming device, or specialized software is required for setup in the field.
Key Specifications
- Enclosure Type: General Purpose
- Enclosure Material: Rugged Anodized Aluminum
- Standard Ranges: 0–10 / 100 / 1,000 / 10,000 ppm and 0–25%
- Optional Ranges: 0–1 / 5 / 10 / 25 / 100%
- Resolution: 0.01 ppm or 0.01%
- Power Configurations: 2-wire loop, 3-wire, or 6-wire
- Analog Outputs: 4–20 mA standard; isolated 4–20 mA, 0–10 VDC, and 0–1 VDC available depending on configuration
- Digital Communication: Optional bi-directional Modbus RS485 ASCII
Flexible Power and Output Configurations
The OMD-507 is available with three electrical configurations to accommodate a wide range of industrial installations:
- 2-Wire Loop Power: Non-isolated 4–20 mA output
- 3-Wire Configuration: Isolated 4–20 mA output
- 6-Wire 12–24 VDC: Non-isolated 4–20 mA, 0–10 VDC, and 0–1 VDC outputs with optional Modbus RS485 ASCII communication
This flexibility allows the OMD-507 to integrate easily into existing process control systems, PLCs, data acquisition systems, and other industrial instrumentation.
Precision Electrochemical Oxygen Sensor
At the heart of the OMD-507 is Southland Sensing’s galvanic electrochemical fuel-cell oxygen sensor, manufactured in-house under strict quality procedures. The self-contained sensor requires minimal maintenance—there are no electrodes to clean and no electrolyte to replenish.
Our precision oxygen sensors are engineered to provide excellent accuracy, stability, response, and service life across a broad range of industrial gas applications.
Trace, Percent and Purity Oxygen Measurement
The OMD-507 can be configured for trace (ppm), percent, or oxygen purity measurement, with measurement ranges extending from 0–10 ppm through 0–100% O₂.
Sensor options are available for different measurement ranges and background gases, including specialized sensors designed for applications containing CO₂ or acidic gas components.
Sensor Housing Options
Multiple sensor housing configurations allow the OMD-507 to be adapted to the installation:
- KF-40 / NW40F direct-fit configuration for inert glovebox applications
- KF-40L direct-fit configuration for customers needing an extended KF flange
- Flow-through sensor housing for continuous gas sampling
- Available process connections include 6 mm, 1/8-inch, and 1/4-inch compression tube fittings
The combination of flexible measurement ranges, configurable electronics, multiple output options, and field-replaceable oxygen sensors makes the OMD-507 well suited for a wide variety of industrial oxygen measurement applications.




Comentários
Ainda não existem comentários.