OMD-675 Oxygen Analyzer
Online Trace Oxygen Analyzer w/ Integral Sample System
The OMD-675 is designed for continuous ultra-trace oxygen measurement in the sub-1 PPM region. The analyzer combines advanced electronics with Southland Sensing’s precision electrochemical oxygen sensor technology in a standard 19″ rack-mount configuration, providing reliable and cost-effective PPB and PPM oxygen measurement with an intuitive, easy-to-use interface.
The analyzer features a 0–1 PPM O₂ full-scale low range with 0.001 PPM (1 PPB) resolution. Additional measurement ranges include 0–10 PPM, 0–100 PPM, 0–1,000 PPM, and 0–25% O₂. The analyzer can be operated in either automatic-ranging or manual-range mode.
Bi-directional MODBUS RS485 RTU communications are available, allowing the OMD-675 to be integrated with process control, data acquisition, and monitoring systems.
Alarm functionality includes two fully adjustable Form C, non-latching relay contacts. Each alarm can be independently configured for high or low oxygen conditions, normally open or normally closed operation, and optional alarm delay. A dedicated Form C, non-latching power-failure alarm relay is also provided as standard.
The OMD-675 utilizes Southland Sensing’s galvanic electrochemical oxygen sensor technology. All oxygen sensors are manufactured in-house under a controlled quality program to provide consistent performance, accuracy, and long-term stability.
Two sensor chemistries are available to accommodate different process gas compositions. The standard TO2-133 sensor is designed for applications containing background gases such as hydrogen, helium, and hydrocarbons. For applications containing acid gases, including carbon dioxide and many natural gas streams, the optional TO2-233 provides an acid-resistant alternative.
Southland Sensing oxygen sensors are completely self-contained and require minimal routine maintenance, with no electrodes to clean or electrolyte to replenish.
Applications :
- Air Liquefaction and Separation
- Pure, Gaseous Hydrocarbon Stream Monitoring
- Gas Purity Certification
- Process Monitoring of Gaseous Monomers
- Fabrication de semi-conducteurs
- Protective Atmosphere Blanketing of Feedstock
- 0–1 PPM O₂ Low Range with 0.001 PPM (1 PPB) Resolution
- Integral Sample System with Bypass / Isolation Valve
- Bi-Directional MODBUS RS485 RTU Communications
- Comprehensive Analog & Digital Outputs
- Standard 19" Rack-Mount Configuration
- Technologie de capteur électrochimique d’oxygène de précision
Spécifications :
- Précision : < ±1 % de la pleine échelle
- Alarms: (2) Adjustable Relay Contacts (1) Power Fail Relay Contact
- Analysis Range: 0-1/10/100/1000ppm/25%
- Étalonnage : périodique
- Communication: Bi-Directional -- MODBUS RS485 RTU
- Dimensions: 19.0 x 5.125 x 7.75 inch
- Affichage : grand écran rétroéclairé
- Boîtier : acier inoxydable brossé
- Flow Sensitivity: 0.5 - 5 SCFH
- Raccord gaz : tube à compression 1/8 pouce
- Output (Analog): 0 - 1V DC, 0 - 10V DC -- Isolated 4 - 20mA
- Alimentation : 100–240 V CA
- Pressure: Inlet, 0.1 - 50 PSIG
- Range ID: 4 - 20 mA analog output -- 0 - 1 V DC analog output
- Temps de réponse : T90 en 10 secondes
- Sample System: Flow Control, 4-way sample / Bypass Valve, Flow Indicator
- Sensor: TO2-133 Trace O2 Sensor
- Sensor Life: 20 - 25 months
- Température : 0–50 °C
- Temperature Compensation: Digital
- Garantie : 12 mois pour analyseur et capteur
- Poids : 13,0 livres
Configurations optionnelles :
- Panel Mount Oxygen Configuration (OMD-677)
- TO2-233 Oxygen Sensor (for use in CO2)