OMD-501X Oxygen Analyzer

Online Oxygen Analyzer, 1/4 DIN Package, Remote Sensor Housing

应用领域

技术规格

OMD-501X Oxygen Analyzer | Product Specifications

Oxygen Analyzer:

The model OMD-501X oxygen analyzer combines a rugged in-line design with SSO2’s precision oxygen sensors. The result is a highly reliable and cost effective compact design with easy-to-use user interface.

Trace ranges include 0 – 10 ppm, 0 – 100 ppm, 0 – 1000ppm, 0-10000ppm and 0 25%. Percent ranges include 0 – 1%, 0 – 5%, 0 – 10%, 0 – 25%, and 0 – 100%. The analyzer can be configured for trace (parts per-million), percent, and purity applications by the user by selecting which sensor is in use in the built-in menu.

The analog output can be manual range selected through the on board menu or the user can take advantage of the auto-range feature using the RANGE ID output allowing easy interface with a PLC, DCS or other control system.

Gas connections are made with compression tube fittings or a direct fit KF-40 Housing.

电源要求

Input Power: 10 – 28 V DC
Input Power: 100 – 240 V AC
Current Draw: 50 mA

Oxygen Sensor Technology:

The oxygen sensors used in the OMD-501X are based on the galvanic electrochemical fuel cell principal. All oxygen sensors are manufactured in house by Southland Sensing Ltd. under a strict quality program.

The standard cells are unaffected by other background gases such as H2, He or Hydrocarbons. The acidic cells work well when acid gases such as CO2 or Natural Gas are present.

传感器采用一体化设计,几乎无需维护,无需清洁电极或添加电解液。

The SSO2 precision oxygen sensors offer excellent performance, accuracy and stability while maximizing the expected life.

Oxygen Sensors:

TO2-1x PPM Oxygen Sensor: Trace Analysis, Standard
TO2-2x PPM Oxygen Sensor: Trace Analysis, Acidic
PO2-160 Percent Oxygen Sensor: Percent Analysis, Standard
PO2-24 Percent Oxygen Sensor: Percent Analysis, Acidic
PO2-1120 Purity Oxygen Sensor: Purity Analysis
TO2-19 Hybrid Oxygen Sensor: Percent or Trace Analysis

Oxygen sensors should be periodically calibrated. Factory recommendation is every 2 – 3 months or as the application dictates. Sensors offer excellent linearity with an air calibration, or calibrate to a certified span gas to maximize accuracy.